Development of a miniature scintillation camera using an NaI(Tl) scintillator and PSPMT for scintimammography
- 23 October 2000
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 45 (11) , 3481-3488
- https://doi.org/10.1088/0031-9155/45/11/326
Abstract
We have developed a small scintillation camera dedicated to breast imaging and have evaluated the performance of the system. In order to increase the limited field of view (FOV) determined by the size of a position-sensitive photomultiplier tube (PSPMT), the imaging characteristics of a diverging-hole collimator (DHC) were also investigated. The small scintillation camera system consists of an NaI(Tl) crystal (60 mm × 60 mm × 6 mm) coupled to a Hamamatsu R3941 PSPMT, a resistor chain circuit, preamplifiers, nuclear instrument modules, an analogue to digital converter and a PC for control and display. The intrinsic energy resolution of the system was 12.9% FWHM at 140 keV. The spatial resolution was measured using a line-slit mask and 99mTc point sources and was 3.1 mm FWHM. The intrinsic sensitivity of the system was approximately 162 counts/s kBq-1. The DHC made it possible to image a larger FOV (75×75 mm2 at the surface of collimator) than a parallel-hole collimator (60×60 mm2). The system sensitivity obtained using the DHC gradually decreased with distance (3% at 1 cm, 6% at 2 cm and 9% at 3 cm). The results demonstrate that the system developed in this study could be utilized clinically to image malignant breast tumours. A DHC can be employed to expand the FOV of the system confined by the size of PSPMT with a modest compromise in the performance of the system.Keywords
This publication has 14 references indexed in Scilit:
- Sensitivity and Specificity of Planar Scintimammography with 99mTc-SestamibiActa Radiologica, 1999
- 99Tcm-MIBI scintimammography for the detection of breast malignanciesNuclear Medicine Communications, 1999
- Delayed Tc-99m Citrate Scintigraphy in the Evaluation of Palpable Breast MassesClinical Nuclear Medicine, 1999
- 99Tcm-MIBI scintimammography in the evaluation of breast lesions and axillary involvementNuclear Medicine Communications, 1999
- Cancer statistics, 1998CA: A Cancer Journal for Clinicians, 1998
- PSPMT and photodiode designs of a small scintillation camera for imaging malignant breast tumorsIEEE Transactions on Nuclear Science, 1997
- A high energy gamma camera using a multiple hole collimator and PSPMTNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1994
- Simulation of imaging with sodium iodide crystals and position-sensitive photomultiplier tubesIEEE Transactions on Nuclear Science, 1994
- Sensitivity and specificity of first screen mammography in the Canadian National Breast Screening Study: a preliminary report from five centers.Radiology, 1986
- Scintillation CameraReview of Scientific Instruments, 1958