Electrophysiological Properties of a New Permanent Endocardial Lead for Uni‐ and Bipolar Pacing
- 1 March 1982
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 5 (2) , 268-274
- https://doi.org/10.1111/j.1540-8159.1982.tb02224.x
Abstract
Unipolar and bipolar electrode systems were compared for electrogram amplitudes and slew rates, signal source impedance, and myocardial stimulation threshold and resistance in 15 consecutive patients who received a new endocardial electrode (Cordis 325-161). The bipolar electrograms showed the highest amplitude in nine of the patients (60%). The unipolar and bipolar electrograms were equal in four patients (26.7%), whereas the unipolar electrograms were highest in only two patients (13.3%). The difference in mean amplitude between bipolar (11.1 mV) and unipolar (10.1 mV) electrograms was statistically significant (p 0.05). Mean slew rates were almost equal (1.7 versus 1.6 V/s; p greater than 0.1). The bipolar electrode system always gave somewhat higher signal source impedance than the unipolar system (p 0.001). The current threshold was significantly lower during bipolar pacing (0.59 mA) in constant current pacing mode, than during unipolar pacing (0.65 mA) (p less than 0.05). No significant differences were found during constant voltage pacing. Stimulation resistance was highest in the bipolar electrode system (p less than 0.001). We conclude that the bipolar electrode system is as good as, or better than, the unipolar system both for ventricular sensing and for pacing.Keywords
This publication has 11 references indexed in Scilit:
- Myopotential Inhibition of Unipolar QRS‐inhibited (VVI) Pacemakers, Assessed by Ambulatory Holter Monitoring of the ElectrocardiogramPacing and Clinical Electrophysiology, 1980
- Signal source impedance of implanted pacemaker electrodes estimated from the spectral ratio between loaded and unloaded electrograms in manMedical & Biological Engineering & Computing, 1980
- Demand Failures Occurring during Permanent Pacing in Patients with Serious Heart DiseasePacing and Clinical Electrophysiology, 1980
- The Interdependence Between Electrogram, Total Electrode Impedance and Pacemaker Input Impedance Necessary to Obtain Adequate Functioning of Demand PacemakersPacing and Clinical Electrophysiology, 1979
- A comparison of unipolar and bipolar electrograms for cardiac pacemaker sensing.Circulation, 1977
- Cardiac pacing and pacemakers IV. Threshold of cardiac stimulationAmerican Heart Journal, 1977
- The ventricular endocardial electrogram and pacemaker sensingThe Journal of Thoracic and Cardiovascular Surgery, 1977
- Interference effect of myopotenitals on function of unipolar demand pacemakers.Heart, 1974
- Failure of Demand Function in Temporary Epicardial Bipolar Pacemaker SystemsThe Annals of Thoracic Surgery, 1973
- Failure of demand pacemaker from low-voltage bipolar ventricular electrogramsPublished by American Medical Association (AMA) ,1971