Self-Referencing Ceramic-Based Multisite Microelectrodes for the Detection and Elimination of Interferences from the Measurement of l-Glutamate and Other Analytes
- 26 January 2001
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 73 (5) , 1037-1042
- https://doi.org/10.1021/ac0010429
Abstract
A self-referencing technique utilizing two microelectrodes on a ceramic-based multisite array is employed for confirmation and elimination of interferences detected by enzyme-based microelectrodes. The measurement of l-glutamate using glutamate oxidase was the test system; however, other oxidase enzymes such as glucose oxidase can be employed. One recording site was coated with Nafion with l-glutamate oxidase and bovine serum albumin (BSA) cross-linked with glutaraldehyde while the other had Nafion with BSA cross-linked with glutaraldehyde. Differences in the chemistry of the two recording sites allowed for identification and elimination of interfering signals to be removed from the analyte response. The electrode showed low detection limits (LOD = 0.98 ± 0.09 μM, signal-to-noise ratio of 3), fast response times (T90 ∼1 s), and excellent linearity (R2 = 0.999 ± 0.000) over the concentration range of 0−200 μM for calibrations of l-glutamate in vitro. The selectivity and dimensions of the multisite electrode allow in vivo glutamate measurements. This electrode has been applied to in vivo measurements of the clearance of locally applied glutamate and release of glutamate in the prefrontal cortex of anesthetized rats. In addition, a similar approach has been applied to the development of a microelectrode for measures of glucose.Keywords
This publication has 14 references indexed in Scilit:
- Direct Immobilization of Glutamate Dehydrogenase on Optical Fiber Probes for Ultrasensitive Glutamate DetectionAnalytical Chemistry, 1999
- Fabrication and Characterization of Sputtered-Carbon Microelectrode ArraysAnalytical Chemistry, 1996
- In Vivo Brain Glucose Measurements: Differential Normal Pulse Voltammetry with Enzyme-Modified Carbon Fiber MicroelectrodesAnalytical Chemistry, 1996
- Determination of l-glutamate and l-glutamine in pharmaceutical formulations by amperometric l-glutamate oxidase based enzyme sensorsJournal of Pharmaceutical and Biomedical Analysis, 1993
- Ultramicrobiosensors for Monitoring of NeurotransmittersAnnals of the New York Academy of Sciences, 1992
- Measurement of Microscale Patchiness in a Turbulent Aquatic Odor Plume Using a Semiconductor-Based MicroprobeThe Biological Bulletin, 1992
- Regional effects of aging on dopaminergic function in the Fischer-344 ratNeurobiology of Aging, 1992
- Design and in vitro studies of a needle-type glucose sensor for subcutaneous monitoringAnalytical Chemistry, 1991
- Multichannel semiconductor-based electrodes for in vivo electrochemical and electrophysiological studies in rat CNSNeuroscience Letters, 1990
- Nafion-coated electrodes with high selectivity for CNS electrochemistryBrain Research, 1984