Resolving Time Measurements of GM and Gas-Flow Counters Using a Time-to-Amplitude Converter
- 1 November 1972
- journal article
- Published by IOP Publishing in Japanese Journal of Applied Physics
- Vol. 11 (11) , 1686-1692
- https://doi.org/10.1143/jjap.11.1686
Abstract
A slow time-to-amplitude converter is fabricated for the time interval measurements of self-quenching type GM and gas-flow counters. With this converter, the resolving times of counters are analyzed. The resolving time has the peak value at the beginning of the GM-threshold voltage and decreases at the higher supplied voltage. The approximate position of the ion sheath of the normal discharge at which the field strength of a counter recovers to the GM-threshold is also determined by comparing the resolving time with the calculated velocity of the ton sheath. The resolving time is proportional to the pulse amplitude at low supplied voltages and is nearly constant at higher voltages. These phenomena can be explained by the recovery curves of the field strength near the center wire. Another application of the coverter is also suggested.Keywords
This publication has 4 references indexed in Scilit:
- Resolving time measurements of Geiger counters using a time analyzerNuclear Instruments and Methods, 1970
- Zur abhängigkeit der totzeit und der erholungskurve von der impulshöhe bei Geiger-Müller zählrohrenNuclear Instruments and Methods, 1968
- Der einflusz von geometrie und füllung auf die eigenschaften bromgelöschter Geiger-Müller-ZählrohreNuclear Instruments and Methods, 1965
- Recovery time and counting characteristics of a G.M. counter at higher intensitiesNuclear Instruments and Methods, 1965