Determination of some parameters for pion radiobiology studies
- 1 May 1977
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 22 (3) , 466-475
- https://doi.org/10.1088/0031-9155/22/3/007
Abstract
The central axis depth-dose and stopping rate distribution of the SIN [Schweizerisches Institut fuer Nuklearforschung] biomedical pion beam is reported. The pion stopping rate in a thin disc of tissue-equivalent plastic was determined using a counter telescope. The dose rate at the position of this disc target was measured using a specially designed parallel-plate tissue-equivalent ionization chamber. Both dose rate and pion stopping rate were a function of depth for beams of 2 different momenta spread. The energy deposition required per pion stop to fit the measured dose rate curves was calculated and 37-40 MeV. From the stopping rate measurements the depth-dose distribution of pion interaction dose (star-dose) and the dose due to the pion slowing down were evaluated. The maximum star-dose was at a depth of about 2 g cm-2 beyond the maximum of the ionization dose.This publication has 2 references indexed in Scilit:
- III. Der Pionenstrahl (π−) für biologische und medizinische Anwendungen am Schweizerischen Institut für Nuklearforschung (SIN)RöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 1974
- Analysis of energy deposition in water around the site of capture of a negative pion by an oxygen or carbon nucleusPhysics in Medicine & Biology, 1972