Abstract
Radio frequency (RF) phase gradients in the receiver coil field pattern can encode flow velocity information in magnetic resonance (MR) images in the form of phase variations. These phase variations are not readily observed in MR images because they are relatively small compared to phase variations caused by static magnetic field (B0) inhomogeneities, susceptibility variations, and other sources. However, the phase contributions from these other sources are independent of the receiver coil. Therefore, the RF phase gradient encoded flow information can be recovered by subtracting images obtained simultaneously using arrays of independent receiver coils and a multiple channel receiver. This flow velocity information can be extracted retrospectively from standard imaging sequences, including flow-compensated sequences. No additional time is required for the flow study as the flow measurements are made using sequences chosen for optimal imaging, and the images from each coil are obtained simultaneously. Initial results indicate that sufficient sensitivity is obtained to make flow measurements in the range of velocities commonly found in the carotid arteries and other major vessels. In principle, the method works with only two receiver coils. However, additional elements provide additional phase measurements that can be used to increase accuracy, remove ambiguities in flow direction or velocity calculations, and increase the region over which velocity measurements can be accurately made.Keywords
This publication has 5 references indexed in Scilit:
- Arrays of mutually coupled receiver coils: Theory and applicationMagnetic Resonance in Medicine, 1991
- Detection of motion using B1 gradientsMagnetic Resonance in Medicine, 1988
- An introduction to NMR imaging: From the Bloch equation to the imaging equationProceedings of the IEEE, 1983
- A flow velocity zeugmatographic interlace for NMR imaging in humansMagnetic Resonance Imaging, 1982