The Implantable Cardioverter‐Defibrillator: Clinical Results
- 1 February 1991
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 14 (2) , 297-301
- https://doi.org/10.1111/j.1540-8159.1991.tb05110.x
Abstract
To evaluate the effectiveness of the automatic implantable cardioverter‐defibrillator (AICD), a 7‐year experience, from 1983–1990, was reviewed. A total of 111 patients received an AICD device. Their ages ranged between 8 and 83 years. Mean age was 63.9 years. There were 91 men and 20 women. Eighty of the patients received the AICD following an out‐of‐hospital cardiac arrest, white 32 were suffering from intermittent symptomatic ventricular tachycardia. The underlying etiology in 97 patients (87%) was ischemic coronary artery disease, in 11 patients (10%) dilated cardiomyopathy, and in 3 patients (3%) idiopathic ventricular fibrillation. Mean ejection fraction was 33.2%. Implantation of the AICD was performed via a left thoracotomy in 39 patients, median sternotomy in 49 patients and subxiphoidsubcostal approach in 23 patients. In‐hospital mortality occurred in one patient who suffered an acute myocardial infarction 4 hours postoperatively. Out‐of‐hospital mortality was observed in 19 patients. There were two arrhythmic deaths. Follow‐up was available for 107 patients. Mean follow‐up was 33.1 months. Sixty‐six patients (62%) had AICD shocks. The initial appropriate shocks occurred during the first postimplantation year in 91% of the patients. In 53 of the survivors, initial AICD shocks took place within 4.4 ± 4.7 months from implantation. Thirteen of the 20 patients who died had received appropriate AICD shocks. In these patients, the time between implantation and first shock was 2.7 ± 3.6 months whereas the time between implantation and death was 11.3 ± 10.3 months (NS). We conclude that the AICD is effective in converting ventricular tachyarrhythmias and prolongs survival.Keywords
This publication has 12 references indexed in Scilit:
- Time to first shock and clinical outcome in patients receiving an automatic implantable cardioverter-defibrillatorJournal of the American College of Cardiology, 1989
- Automatic Implantable Cardioverter-Defibrillator: Techniques of Implantation and ResultsThe Annals of Thoracic Surgery, 1986
- The automatic implantable cardioverter-defibrillator: An overviewJournal of the American College of Cardiology, 1985
- Clinical experience, complications, and survival in 70 patients with the automatic implantable cardioverter/defibrillator.Circulation, 1985
- Identification of patients at high risk for sudden cardiac deathThe American Journal of Cardiology, 1984
- Surgical Techniques for Implanting the Automatic Implantable DefibrillatorPacing and Clinical Electrophysiology, 1984
- Clinical Performance of the Implantable Cardioverter‐DefibrillatorPacing and Clinical Electrophysiology, 1984
- Perspectives on the current treatment of cardiac arrhythmiasThe American Journal of Cardiology, 1984
- Termination of Malignant Ventricular Arrhythmias with an Implanted Automatic Defibrillator in Human BeingsNew England Journal of Medicine, 1980
- Standby automatic defibrillator. An approach to prevention of sudden coronary deathArchives of internal medicine (1960), 1970