Verification of Multichannel Liquid Dispenser Performance in the 4–30 μL Range by Using Optical Pathlength Measurements in Microplates

Abstract
Pipettors and liquid handling systems should be checked regularly to verify their precision and accuracy. Professional organizations recommend performing 4-replicate tests monthly and 10-replicate tests at least quarterly (1)(2)(3)(4). Traditional gravimetric procedures are adequate to meet these testing requirements for single-channel pipettors but are tedious and impractical for multichannel devices. An alternative is to dispense a solution of colored dye into wells of a microplate and to measure the resulting absorbance values in a microplate reader. Volume calibration is performed with a calibration curve relating absorbance to volume (5).