Verification of Multichannel Liquid Dispenser Performance in the 4–30 μL Range by Using Optical Pathlength Measurements in Microplates
Open Access
- 1 October 1998
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 44 (10) , 2206-2208
- https://doi.org/10.1093/clinchem/44.10.2206
Abstract
Pipettors and liquid handling systems should be checked regularly to verify their precision and accuracy. Professional organizations recommend performing 4-replicate tests monthly and 10-replicate tests at least quarterly (1)(2)(3)(4). Traditional gravimetric procedures are adequate to meet these testing requirements for single-channel pipettors but are tedious and impractical for multichannel devices. An alternative is to dispense a solution of colored dye into wells of a microplate and to measure the resulting absorbance values in a microplate reader. Volume calibration is performed with a calibration curve relating absorbance to volume (5).Keywords