A Multicentre European Evaluation of the Kodak EKTACHEM GLU/BUN Analyzer Using NCCLS Guidelines and Other Approaches
- 1 January 1982
- journal article
- research article
- Published by Walter de Gruyter GmbH in cclm
- Vol. 20 (4) , 207-216
- https://doi.org/10.1515/cclm.1982.20.4.207
Abstract
This paper describes the field performance of the Kodak EKTACHEM GLU/BUN [glucose/blood urea nitrogen] Analyzer for glucose and urea. NCCLS [National Committee for clinical Laboratory standards] protocols PSEP-2, 3 and 4 were used which enable manufacturers to establish performance claims concerning the precision and accuracy of an analytical system. This multicenter trial used 4 analysers in 4 European countries, United Kingdom, France, West Germany and Italy to assess within and between laboratory performances. Freeze dried control materials were used for the performance check experiment PSEP-2 and for the replication experiment PSEP-3. The replication study, although very time consuming, was straight forward to undertake and the results were comparable between centres. Compliance with the protocol for comparison of methods experiment PSEP-4 in which laboratories used their own hospital patient samples was more difficult. The problems with obtaining suitable samples and the performance of comparative methods are discussed in detail.This publication has 6 references indexed in Scilit:
- DuPont aca III performance as tested according to NCCLS guidelines.Clinical Chemistry, 1979
- Characteristics of statistical parameters used to interpret least-squares results.Clinical Chemistry, 1978
- Applications of statistics in clinical chemistry: A critical evaluation of regression linesClinica Chimica Acta; International Journal of Clinical Chemistry, 1975