Twenty Years Experience in Fracture Healing Measurement with Strain Gauges
- 1 December 1984
- journal article
- Published by SLACK, Inc.
- Vol. 7 (12) , 1823-1826
- https://doi.org/10.3928/0147-7447-19841201-08
Abstract
The external fixator provides a mechanical connection between the bone fragments and recording devices or power systems. We have used this characteristic of external fixation since 1965 to record deformations occurring at the fracture site and to measure the mechanical characteristics of fracture healing in more than 500 patients. As in all osteosyntheses, the fixation material (here the external rod) resists the mechanical forces applied to the bone fragments. As healing progresses, the bone participates more and more in the weight bearing of the system and the rod deforms less. Seven types of deformation curves have been defined, characterizing seven different modes of healing: 1) normal healing, 2) slow healing, 3) delayed union, 4) arrest in evolution, 5) pseudarthrosis, 6) resorption of the callus, and 7) breakage of the callus.Keywords
This publication has 6 references indexed in Scilit:
- Serial Strain Gauge Measurement of Bone Healing in Hoffmann® External FixationPublished by SLACK, Inc. ,1984
- Analysis of Single-Sided External Fracture FixationEngineering in Medicine, 1979
- Measurement of the stiffness of fracture callus in vivo. A theoretical studyJournal of Biomechanics, 1972
- Measurements of Stability of Crural Fractures Treated with Hoffmann Osteotaxis: 2.Measurements on Crural FracturesActa Orthopaedica, 1972
- Measurement of Stability of Tibial Fractures: A Mechanical MethodActa Orthopaedica, 1970
- BIOMECHANICS IN ORTHOPAEDIC SURGERYPublished by Elsevier ,1965