Evaluating Microkeratome Efficacy by 3D Corneal Lamellar Flap Thickness Accuracy and Reproducibility Using Artemis VHF Digital Ultrasound Arc-scanning
- 1 May 2006
- journal article
- research article
- Published by SLACK, Inc. in Journal of Refractive Surgery
- Vol. 22 (5) , 431-440
- https://doi.org/10.3928/1081-597x-20060501-03
Abstract
To present a method that determines the efficacy of flap creation using a microkeratome. Thirty-six eyes of 18 patients were included in the study, operated by two experienced surgeons using the Moria LSK-One microkeratome. Very high-frequency (VHF) digital ultrasound arc-scanning using Artemis technology was used to measure the Reinstein Flap Thickness Profile created by addition of the preoperative epithelial thickness profile to the postoperative stromal component of the flap. Descriptive statistics revealed central flap thickness accuracy and reproducibility. Univariate and multivariate regression analysis was used to determine correlations between central flap thickness and preoperative clinical parameters. Three-dimensional flap thickness maps demonstrated the mean, median, standard deviation, and range of thickness. Intended flap thickness was 160 microm. Mean central flap thickness was 161 microm in the right eye and 166 microm in the left eye. Central reproducibility (standard deviation of the mean) was 30.3 microm, with a range of 106 to 228 microm. Central flap thickness correlated to preoperative corneal thickness measured by VHF digital ultrasound but not by Orbscan. On qualitative analysis, no consistency among eyes was observed in the three-dimensional flap thickness map. Flaps were on average thinner in the center, but not in all cases. A systematic qualitative and quantitative statistical evaluation of microkeratome accuracy and reproducibility can be performed by measuring corneal flap thickness with arc-scanning VHF digital ultrasound.Keywords
This publication has 15 references indexed in Scilit:
- Accuracy and Precision of the Amadeus Microkeratome in Producing LASIK FlapsCornea, 2003
- Factors That Affect Corneal Flap Thickness With the Hansatome MicrokeratomeJournal of Refractive Surgery, 2002
- Arc-scanning Very High-frequency Digital Ultrasound for 3D Pachymetric Mapping of the Corneal Epithelium and Stroma in Laser in situ KeratomileusisJournal of Refractive Surgery, 2000
- Central corneal pachymetry in patients undergoing laser in situ keratomileusis11The authors have no proprietary or financial interest in any product mentioned in this article.Ophthalmology, 1999
- Reproducibility of Corneal Flap Thickness in LASIKOphthalmic Surgery, Lasers and Imaging Retina, 1999
- Very high-frequency ultrasound corneal analysis identifies anatomic correlates of optical complications of lamellar refractive surgeryOphthalmology, 1999
- Visualizing VHF ultrasound of the human corneaIEEE Computer Graphics and Applications, 1999
- Progressive Keratektasie nach Laser-in-situ-Keratomileusis (LASIK)Klinische Monatsblätter für Augenheilkunde, 1998
- Epithelial and Corneal Thickness Measurements by High-frequency Ultrasound Digital Signal ProcessingOphthalmology, 1994
- Comparison of Central Corneal Thickness Measurements between Optical and Ultrasound PachometersOptometry and Vision Science, 1992