Effective wedge angles for 6-MV wedges

Abstract
Measurements were undertaken with 30° and 45° large wedges on a 6‐MV linac to determine the effective wedge angle for various combinations of open and wedged fields. The validity of Tatcher's equation, relating effective wedge angle to maximum dose weightings, was examined over a range of field sizes from 5×5 to 20×20 cm. An alternative equation involving only central axis quantities was also investigated. The results obtained from an analysis of point dose measurements indicate that, for these wedges, either equation yields sufficient accuracy for clinical purposes.