Interference Comparator for the Calibration of Line Standards Directly in Terms of Standard Wavelengths
- 1 May 1961
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 32 (5) , 549-554
- https://doi.org/10.1063/1.1717434
Abstract
An interference comparator has been built for the routine calibration of line standards or scales by direct reference to the vacuum wavelength of the 0.606 μ line of Kr86 which has recently become the international standard of length. The principles of construction and the performance of the comparator are given; it is shown to yield an accuracy of calibration which is mostly limited by the nature of present-day line standards or scales, viz., a few hundredths of a micron in lengths up to one meter.Keywords
This publication has 2 references indexed in Scilit:
- Obliquity Effects in InterferometryOptica Acta: International Journal of Optics, 1957