Input impedance of radiocephalic arteriovenous fistulae for haemodialysis access: its value in predicting early failure
- 1 November 1994
- journal article
- clinical trial
- Published by IOP Publishing in Physiological Measurement
- Vol. 15 (4) , 469-480
- https://doi.org/10.1088/0967-3334/15/4/008
Abstract
This study is concerned with an assessment of the quality of the blood vessels used in the construction of radiocephalic arteriovenous fistulae for haemodialysis vascular access in 20 patients. Following non-invasive preoperative assessment of the cephalic vein by means of a colour Doppler scanner, input impedance was determined intraoperatively from blood flow and blood pressure measured about 2 cm downstream of the anastomosis. The patients were re-assessed 1 day and 2, 4, 6 and 12 weeks after surgery, and the diameter and the flow through the fistulae, the location of major branches, and the presence and severity of stenoses in the cephalic veins were determined non-invasively. Five fistulae failed within the 12 weeks period following surgery, mainly due to thrombosis. In a further 3 patients, the fistulae were patent but did not achieve an adequately high blood flow and dilate sufficiently for haemodialysis and were therefore considered to be essentially failed. The impedance moduli from 0 to 10 Hz as well as the average impedance modulus of successful fistulae were significantly lower than those of fistulae that subsequently failed. An attempt was made to locate the presence of reflection sites (i.e. stenoses and branches) from the minima in the impedance modulus spectra. Their locations were compared with those of the reflection sites detected by ultrasound 2 weeks after surgery. However, not all reflection sites detected 2 weeks after surgery were located by the impedance method. It is recommended that both preoperative examination and intraoperative haemodynamic measurements be made to improve the accuracy of the assessment.Keywords
This publication has 13 references indexed in Scilit:
- Quantitation of vascular outflow by measurement of impedanceJournal of Vascular Surgery, 1991
- A pulse-foot-seeking algorithm for Doppler ultrasound waveformsClinical Physics and Physiological Measurement, 1988
- Pulse-generated runoff: a new method of determining calf vessel patencyBritish Journal of Surgery, 1988
- Vascular Access for Hemodialysis Patency Rates and Results of RevisionAnnals of Surgery, 1985
- The hydraulic impedance of the lower limb: Its relevance to the success of bypass operations for occlusion of the superficial femoral arteryBritish Journal of Surgery, 1976
- Effective length of the arterial systemAnnals of Biomedical Engineering, 1975
- The Transmission Characteristics of Large and Small Pressure Waves in the Abdominal vena CavaIEEE Transactions on Biomedical Engineering, 1969
- Input Impedance of the Systemic CirculationCirculation Research, 1967
- Chronic Hemodialysis Using Venipuncture and a Surgically Created Arteriovenous FistulaNew England Journal of Medicine, 1966
- An Approach to an Analysis of the Arterial Pulse Wave I. Oscillations in an Attenuating LinePhysics in Medicine & Biology, 1957