Flow Injection Analysis of Glucose in Human Serum by Chemiluminescence
- 1 January 1982
- journal article
- research article
- Published by Taylor & Francis in Analytical Letters
- Vol. 15 (21-22) , 1751-1766
- https://doi.org/10.1080/00032718208065151
Abstract
Combining the rigidly controllable conditions of flow injection analysis with detection by chemiluminescence, a method for determination of glucose in serum via enzymatic degradation by glucose oxidase is described. With a SD of less than 2% the sampling frequency is 75 samples/h. Results obtained by the proposed procedure are compared with those acquired by a routine AutoAnalyzer method used at a local hospital. Any influence on the yield of chemiluminescence due to difference in viscosity of the injected samples can be effectively eliminated.This publication has 6 references indexed in Scilit: