An investigation of tomotherapy beam delivery
- 1 March 1997
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 24 (3) , 425-436
- https://doi.org/10.1118/1.597909
Abstract
Experimental simulations for tomotherapy beam delivery were performed using a computer-controlled phantom positioner, a cylindrical phantom, and a 6 MV x-ray slit beam. Both continuous helical beam and sequential segmented tomotherapy (SST) beam deliveries were evaluated. Beam junctioning problem due to couch indexing error or field width errors presented severe dose uniformity perturbations for SST, while the problem was minimized for helical beam delivery. Longitudinal breathing motions were experimentally simulated for helical and SST beam delivery. While motions reduced the dose uniformity perturbations for SST, small artifacts in dose uniformity can be introduced for helical beam delivery. With typical breath frequency and magnitude, for a slit beam of 2.0 cm width at 4 rpm, the dose uniformity perturbation was not significant. A running start/stop technique was implemented with helical beam delivery to sharpen the 20%-80% longitudinal dose fall-off from 1.5 to 0.5 cm. The latter was comparable to the corresponding dose penumbra of a conventional 6 MV 10 x 10 cm2 field. All together, helical beam delivery showed advantages over SST for tomotherapy beam delivery under similar delivery conditions.Keywords
This publication has 16 references indexed in Scilit:
- A filtered backprojection dose calculation method for inverse treatment planningMedical Physics, 1994
- Optimization by simulated annealing of three-dimensional, conformal treatment planning for radiation fields defined by a multileaf collimator: II. Inclusion of two-dimensional modulation of the X-ray intensityPhysics in Medicine & Biology, 1992
- Optimization of conformal radiotherapy dose distributions by simulated annealing: II. Inclusion of scatter in the 2D techniquePhysics in Medicine & Biology, 1991
- A unified approach to the optimization of brachytherapy and external beam dosimetryInternational Journal of Radiation Oncology*Biology*Physics, 1991
- Methods of image reconstruction from projections applied to conformation radiotherapyPhysics in Medicine & Biology, 1990
- The verification of an inverse problem in radiation therapyInternational Journal of Radiation Oncology*Biology*Physics, 1990
- An inverse problem in radiation therapyInternational Journal of Radiation Oncology*Biology*Physics, 1990
- Optimization of stationary and moving beam radiation therapy techniquesRadiotherapy and Oncology, 1988
- A problem in rotation therapy with X RaysInternational Journal of Radiation Oncology*Biology*Physics, 1987
- Solution of an integral equation encountered in rotation therapyPhysics in Medicine & Biology, 1982