A fast coherent synchrotron radiation monitor for the bunch length of the short CEBAF bunches
- 1 January 1996
- proceedings article
- Published by AIP Publishing in AIP Conference Proceedings
- Vol. 367 (1) , 502-511
- https://doi.org/10.1063/1.50289
Abstract
A novel bunch length monitor for short (down to subpicosecond) electron bunches has been developed in a collaboration between CEBAF and the University of Virginia (UVA), using coherent synchrotron radiation (CSR) detection techniques. The monitor employs a state of the art ‘‘narrowband’’ GaAs Schottky whisker diode developed by the UVA group, and has the following features: it is non‐invasive, compact, and low cost, it has fast rise time, low noise, high sensitivity, and it operates at room temperature. In this paper, the design parameters and performance of the monitor and selected measurement results will be presented.Keywords