Evaluation of the Beckman "System TR Enzyme Analyzer"
- 1 July 1975
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 21 (8) , 1107-1112
- https://doi.org/10.1093/clinchem/21.8.1107
Abstract
We evaluated 16 claims made by Beckman Instruments, Inc. for its Enzyme Analyzer (System TR), under a rigid written protocol for the Product Evaluation Subcommittee of the Standards Committee of the College of American Pathologists. We found the following to be within the company's specifications: (a) accuracy and precision of the temperature control; (b) accuracy and precision of the sample and reagent pipets; (c) instrument precision, both within-run and between-day; (d) carry-over from a sample with activity > 1000 U/liter; (e) instrument-to-instrument variation; (f) analytical linearity; (g) analysis time; (h) correlation of the instrumentprinted answer with the activity calculated manually from a strip-chart recorder; (i) precision of the instrument's built-in electronic "standard"; (j) effectiveness of the over-range indicators; and (k) correlation between results of these enzyme assay methods and those for kinetic methods used in our laboratory. The instrument performed well.Keywords