Temperature measurement under high pressure using commercial platinum resistors
- 1 March 1979
- journal article
- Published by IOP Publishing in Journal of Physics E: Scientific Instruments
- Vol. 12 (3) , 185-186
- https://doi.org/10.1088/0022-3735/12/3/011
Abstract
Platinum resistance thermometry has been used for temperature measurement in situ in a high-pressure vessel up to 360 MPa. Commercially available 100 Omega platinum resistors have been calibrated against pressure for temperatures between -60 and +100 degrees C. A convenient small four-wire electrical connector has been developed for use up to 500 MPa.Keywords
This publication has 3 references indexed in Scilit:
- Glass enclosed resistance thermometer for temperature measurement inside a high-pressure vesselReview of Scientific Instruments, 1978
- Large thermocouple thermometry errors caused by magnetic fieldsReview of Scientific Instruments, 1977
- Temperature Measurement under PressurePublished by Elsevier ,1975