Multivariate analysis of TLD orientation effects
- 1 July 1980
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 7 (4) , 352-354
- https://doi.org/10.1118/1.594707
Abstract
The effect of orientation on extruded thermoluminescent dosimeters was investigated. TLD [thermoluminescent dosimeter] placed on the surface and within a phantom were exposed separately to 5 diagnostic beam qualities and to 60Co .gamma. rays. The response of dosimeters on the phantom surface varied with orientation and was energy-dependent. In the phantom and with 60Co, no orientation effects were observed.Keywords
This publication has 1 reference indexed in Scilit:
- Effects of total dose on energy dependence of TLD-Ioo LiF dosimetersThe International Journal of Applied Radiation and Isotopes, 1968