Cryoballoon Pulmonary Vein Isolation Guided by Transesophageal Echocardiography: Novel Aspects on an Emerging Ablation Technique
- 30 October 2009
- journal article
- clinical trial
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 20 (11) , 1197-1202
- https://doi.org/10.1111/j.1540-8167.2009.01524.x
Abstract
Background: Pulmonary vein (PV) isolation using a balloon‐mounted cryoablation system is a new technology for the percutaneous treatment of atrial fibrillation (AF). Transesophageal echocardiography (TEE) allows real‐time visualization of cryoballoon positioning and successful vein occlusion via color Doppler. We hypothesized that PV mechanical occlusion monitored with TEE could predict effective electrical isolation. Methods: We studied 124 PVs in 30 patients. Under continuous TEE assessment, a cryoballoon was placed in the antrum of each PV aiming for complete PV occlusion as documented by color Doppler. At the end of the procedure, PV electrical isolation was evaluated using a circumferential mapping catheter. Results: Of the 124 PVs studied, 123 (99.2%) could be visualized by TEE: the antrum was completely visualized in 80 of them (64.5%), partially in 36 (29.0%), and only disappearance of proximal flow could be observed in the remaining 7 PVs (5.7%). Vein occlusion could be achieved in 111 of the 123 (90.2%) visualized PVs. Postinterventional mapping demonstrated electrical isolation in 109 of 111 occluded PVs (positive predictive value 98.2%) and only in 1 of 12 nonoccluded PVs (negative predictive value 91.7%, P < 0.001). After a mean follow‐up of 7.4 ± 3.7 months, 73.3% of patients remained in sinus rhythm without antiarrhythmic drugs. Conclusion: Color Doppler documented PV occlusion during cryoballoon ablation can predict effective electrical isolation.Keywords
This publication has 23 references indexed in Scilit:
- The 'single big cryoballoon' technique for acute pulmonary vein isolation in patients with paroxysmal atrial fibrillation: a prospective observational single centre studyEuropean Heart Journal, 2008
- One year follow-up after cryoballoon isolation of the pulmonary veins in patients with paroxysmal atrial fibrillationEP Europace, 2008
- Pulmonary vein isolation predicts freedom from arrhythmia after circumferential antral ablation for paroxysmal atrial fibrillationJournal of Cardiovascular Medicine, 2007
- Electrophysiologic and Anatomic Characterization of Sites Resistant to Electrical Isolation During Circumferential Pulmonary Vein Ablation for Atrial Fibrillation: A Prospective StudyJournal of Cardiovascular Electrophysiology, 2007
- A New Approach for Complete Isolation of the Posterior Left Atrium Including Pulmonary Veins for Atrial FibrillationJournal of Cardiovascular Electrophysiology, 2007
- Small or Large Isolation Areas Around the Pulmonary Veins for the Treatment of Atrial Fibrillation?Circulation, 2007
- Identifying far-field superior vena cava potentials within the right superior pulmonary veinHeart Rhythm, 2006
- Efficacy and Safety of Circumferential Pulmonary Vein Isolation Using a Novel Cryothermal Balloon Ablation SystemJournal of the American College of Cardiology, 2005
- Worldwide Survey on the Methods, Efficacy, and Safety of Catheter Ablation for Human Atrial FibrillationCirculation, 2005
- Atrio-Esophageal Fistula as a Complication of Percutaneous Transcatheter Ablation of Atrial FibrillationCirculation, 2004