Endoleak after stent-graft treatment of abdominal aortic aneurysm: a meta-analysis of clinical studies
- 1 May 1999
- journal article
- research article
- Published by Oxford University Press (OUP) in British Journal of Surgery
- Vol. 86 (5) , 581-587
- https://doi.org/10.1046/j.1365-2168.1999.01119.x
Abstract
Background: Endoleak is the major complication after endovascular treatment of abdominal aortic aneurysm (AAA) and its incidence seems to remain significant. Little is known about the association of device type and configuration with respect to the incidence, location, time of onset and fate of endoleakage. Methods: A meta-analysis was performed via a Medline search of clinical studies after 1995 dealing with the endovascular treatment of AAA. Details of number of patients treated, configuration and type of endovascular device were collected. Data concerning site of origin, time of occurrence and fate of the endoleak were retrieved, along with information on change in diameter of the aneurysm with time. Results: The 23 publications included reported on 1189 patients. The 1118 patients with successfully inserted transfemoral endovascular grafts experienced 270 endoleaks (24 per cent). The majority arose from the distal stent attachment site (36 per cent), were present immediately after stent-graft placement (66 per cent) and were persistent in time (37 per cent). Tube grafts were more frequently affected by endoleakage (35 per cent; P < 0·0001), especially at the distal stent attachment site (51 per cent), than bifurcated grafts (18 per cent; P = 0·004) and aortounilateral devices (20 per cent; P = 0·70). Self- expandable stent-grafts were more frequently associated with endoleaks (25 per cent) than balloon-expandable stent-grafts (17 per cent) (P = 0·037). Conclusion: Endovascular treatment of AAA is an evolving field. Even after the initial learning curve and attention to device-related problems, it is still accompanied by a significant number of endoleaks. Uniform presentation of results of treatment is necessary for analysing the effect of differences between patients, aneurysm morphology and device type.Keywords
This publication has 36 references indexed in Scilit:
- Clinical consequences of periprosthetic leak after endovascular repair of abdominal aortic aneurysmJournal of Vascular Surgery, 1998
- Endovascular treatment of infrarenal abdominal aneurysms by the Stentor system: Preliminary results of 79 casesJournal of Vascular Surgery, 1997
- Endovascular aortic stent-grafts: transcatheter embolization of persistent perigraft leaks.Radiology, 1997
- Reporting standards for infrarenal endovascular abdominal aortic aneurysm repairJournal of Vascular Surgery, 1997
- Reduction in aortic aneurysm size: Early results after endovascular graft placementJournal of Vascular Surgery, 1997
- Endoluminal Stent–Grafts for Infrarenal Abdominal Aortic AneurysmsNew England Journal of Medicine, 1997
- Early results of endovascular aortic aneurysm surgery with aortouniiliac graft, contralateral iliac occlusion, and femorofemoral bypassJournal of Vascular Surgery, 1997
- Preliminary clinical outcome and imaging criterion for endovascular prosthesis development in high-risk patients who have aortoiliac and traumatic arterial lesionsJournal of Vascular Surgery, 1996
- Endovascular Repair of Aortic Aneurysms, Arteriovenous Fistulas, and False AneurysmsWorld Journal of Surgery, 1996
- Delayed rupture of aortic aneurysms following endovascular stent graftingThe American Journal of Surgery, 1995