Failure of ceramic hip endoprostheses by slow crack growth - lifetime prediction
- 13 September 1982
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research
- Vol. 16 (5) , 705-713
- https://doi.org/10.1002/jbm.820160516
Abstract
The slow crack propagation velocity v, which depends on the stress intensity factor K, has been measured for three different high‐density aluminas. Assuming a load history for the prosthesis which approximates conditions during walking, the lifetimes of the components of a total hip endoprothesis were calculated as a function of the initial flaw size. Different geometries for the preexisting cracks, as well as varying physiological parameters, were considered. It is shown that the lifetime is very dependent on the quality of the material and that it is reduced by increased body weight, walking speed, and varus positioning. Comparison of failure behavior in different parts of the femoral component showed that flaws in the stem must be one order of magnitude smaller than those in the neck to achieve comparable lifetimes.Keywords
This publication has 7 references indexed in Scilit:
- Use of fracture mechanics theory in lifetime predictions for alumina and bioglass‐coated aluminaJournal of Biomedical Materials Research, 1979
- Strength of a dense alumina ceramic after aging in vitroJournal of Biomedical Materials Research, 1978
- A comparative experimental study of stresses in femoral total hip replacement components: The effects of prostheses orientation and acrylic fixationJournal of Biomechanics, 1976
- Proof testing of ceramic materials?an analytical basis for failure predictionInternational Journal of Fracture, 1974
- Subcritical Crack Growth in CeramicsPublished by Springer Nature ,1974
- Crack propagation in ceramic materials under cyclic loading conditionsMetallurgical Transactions, 1974
- A simple method for evaluating slow crack growth in brittle materialsInternational Journal of Fracture, 1973