A High Resolution, Four-Terminal, ac Resistance Bridge for Low Temperature Use
- 1 April 1972
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 43 (4) , 612-613
- https://doi.org/10.1063/1.1685704
Abstract
Using operational amplifiers in the follower configuration, a four‐terminal, ac resistance bridge can be inexpensively constructed for use with low temperature resistance thermometers from 10 Ω to 10 kΩ (and higher if necessary). Operating the bridge with 1 mV across the thermometer (providing power dissipation of 1 nW at 1 kΩ), the resolution and stability over several hours are 0.003% of the resistance.Keywords
This publication has 1 reference indexed in Scilit:
- Experimental investigations of critical phenomenaReports on Progress in Physics, 1967