Abstract
Described is an experimental verification technique for cesium brachytherapy line source emission. The dose rate levels produced by a line source algorithm are compared with measurements taken around J-series stainless-steel sheathed cesium sources with a p-type semiconductor diode detector. Further study of isodensity contours is undertaken using a film dosimetery technique. The errors involved in using these techniques are assessed.

This publication has 0 references indexed in Scilit: