Clinical evaluation of the Accutracker II ambulatory blood pressure monitor: assessment of performance in two countries and comparison with sphygmomanometry and intra-arterial blood pressure at rest and during exercise
- 30 November 1989
- journal article
- research article
- Published by Wolters Kluwer Health in Journal Of Hypertension
- Vol. 7 (12) , 967-975
- https://doi.org/10.1097/00004872-198912000-00007
Abstract
In order to assess the Accutracker II (Suntech Medical Instruments, Raleigh, North Carolina, USA) a relatively new ambulatory blood pressure (BP) monitor, versus standard forms of BP measurement, we compared same- and contralateral-arm measurements made, via a t-tube connected to a mercury column sphygmomanometer, by two clinicians using a teaching stethoscope and by intra-arterial recordings. Average systolic BP values obtained using the Accutracker II were similar to both the mercury column and intra-arterial determinations, but average diastolic BP values were lower than both the average mercury column (2.8 .+-. 4.2 mmHg, P < 0.001) and intra-arterial measurements (2.0 .+-. 4.7 mmHg, P < 0.02). During isometric exercise and 100-watt bicycle exercise, there were greater limits of agreement for the differences in BP between the Accutracker II and the intra-arterial transducer than were observed for the resting measurements, but these differences were no greater than those observed between intra-arterial and clinician-determined BP measurements. The clinical performance of the Accutracker II was assessed using 119 hypertensive subjects (84 in Norway and 35 in the USA) who wore the monitor for 24 h. While there was good-to-excellent data return in both countries, there were significantly less error codes secondary to excessive arm motion observed in Norway. Our data demonstrate that the Accutracker II is quite accurate compared with both the mercury column and intra-arterial methods of measuring BP, and performs well during 24 h outpatient activities. Our findings also indicate certain geographical differences which may be important in the performance of ambulatory BP-monitoring studies.Keywords
This publication has 5 references indexed in Scilit:
- Clinical evaluation of the Colin ABPM 630 at rest and during exercise: an ambulatory blood pressure monitor with gas-powered cuff inflationJournal Of Hypertension, 1989
- 24-HOUR BLOOD-PRESSURE MONITORING - EVALUATION OF SPACELABS 5300 MONITOR BY COMPARISON WITH INTRA-ARTERIAL BLOOD-PRESSURE RECORDING IN AMBULANT SUBJECTS1988
- Validation of portable noninvasive blood pressure monitoring devices: Comparisons with intra-arterial and sphygmomanometer measurementsAmerican Heart Journal, 1988
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Aortic‐Radial Pulse Transit Time and ECG Q‐Wave to Radial Pulse Wave Interval as Indices of Beat‐By‐Beat Blood Pressure ChangePsychophysiology, 1983