A new catheter design using needle electrode for subendocardial RF ablation of ventricular muscles: finite element analysis and in vitro experiments
- 1 January 2000
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 47 (1) , 23-31
- https://doi.org/10.1109/10.817616
Abstract
Radio-frequency (RF) cardiac ablation has been very successful for treating arrhythmias related with atrioventricular junction and accessory pathways with successful cure rates of more than 90%. Even though ventricular tachycardia (VT) is a more serious problem, it is known to be rather difficult to cure VT using RF ablation. In order to apply RF ablation to VT, we usually need to create a deeper and wider lesion. Conventional RF ablation electrodes often fail to produce such a lesion. We propose a catheter-electrode design including one or more needle electrodes with a diameter of 0.5-1.0 mm and length of 2.0-10 mm to create a lesion large enough to treat VT. One temperature sensor could be placed at the middle of the needle electrode for temperature-controlled RF ablation. From finite element analyses and in vitro experiments, we found that the depth of a lesion is 1-2 mm deeper than the insertion depth of the needle and the width increases as we increase the diameter of the needle and the time duration. We showed that a single needle electrode can produce a lesion with about 10-mm width and any required depth. If a wider lesion is required, more than one needle with suggested structures can be used. Or, repeated RF ablations around a certain area using one needle could produce a cluster of lesions. In some cases, a catheter with both conventional electrode and needle electrode at its tip may be beneficial to take advantage of both types of electrode.Keywords
This publication has 20 references indexed in Scilit:
- Effects of temperature sensor placement on performance of temperature-controlled ablation [cardiac arrhythmias application]Published by Institute of Electrical and Electronics Engineers (IEEE) ,2002
- Temperature distribution under cooled electrodes during radiofrequency catheter ablationPublished by Institute of Electrical and Electronics Engineers (IEEE) ,2002
- Finite element analyses of uniform current density electrodes for radio-frequency cardiac ablationIEEE Transactions on Biomedical Engineering, 2000
- Transcatheter Subendocardial InfusionCirculation, 1996
- Three-dimensional finite element analysis of current density and temperature distributions during radio-frequency ablationIEEE Transactions on Biomedical Engineering, 1995
- Development of a new intraoperative radiofrequency ablation technique using a needle electrodeThe Annals of Thoracic Surgery, 1994
- Numerical model for radio-frequency ablation of the endocardium and its experimental validationIEEE Transactions on Biomedical Engineering, 1994
- A computer simulation of radio-frequency ablation of the endocardiumIEEE Transactions on Biomedical Engineering, 1994
- Chemical ablation by subendocardial injection of ethanol via catheter -- preliminary results in the pig heartEuropean Heart Journal, 1991
- Radiofrequency Catheter Ablation: The Effect of Electrode Size on Lesion Volume In VivoPacing and Clinical Electrophysiology, 1990