Validation of the SpaceLabs 90207 ambulatory blood pressure monitor for use in pregnancy
- 1 October 1993
- journal article
- Published by Wiley in BJOG: An International Journal of Obstetrics and Gynaecology
- Vol. 100 (10) , 904-908
- https://doi.org/10.1111/j.1471-0528.1993.tb15104.x
Abstract
Objective To determine the accuracy of the SpaceLabs 90207 ambulatory blood pressure monitor in pregnant women. Design Prospective observational study. Setting Queen Charlotte's and Chelsea Hospital for Women. Subjects One hundred and twenty‐two pregnant women from the antenatal population of Queen Charlotte's Hospital. Methods Evaluation was carried out according to the British Hypertension Society (BHS) Protocol. Results The mean difference between observer and device for systolic blood pressure was 3 (SD 4) mmHg. For diastolic blood pressure the difference was 5 (SD 6) mmHg for Phase IV (muffling of Korotkoff sound), and 4 (SD 4) mmHg for Phase V (disappearance of Korotkoff sound). According to the BHS grading criteria, the device reached a B grading for systolic blood pressure. B grading was also reached for diastolic blood pressure (Phase V) but grade C for diastolic blood pressure (Phase IV). The accuracy criteria stipulated by the Association for the Advancement of Medical Instrumentation (AAMI) were also met for both systolic and diastolic blood pressure. Conclusion The SpaceLabs 90207, when compared with trained observers, is accurate in determining both systolic and diastolic blood pressure for clinical purposes, as judged by the BHS Protocol in pregnancy.Keywords
This publication has 6 references indexed in Scilit:
- Accuracy of the SpaceLabs 90207 determined by the British Hypertension Society ProtocolJournal Of Hypertension, 1991
- British Hypertension ProtocolJournal Of Hypertension, 1991
- Blood pressure measurement in pregnancyBJOG: An International Journal of Obstetrics and Gynaecology, 1991
- The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systemsJournal Of Hypertension, 1990
- EFFECTS OF BLOOD-PRESSURE MEASUREMENT BY THE DOCTOR ON PATIENT'S BLOOD PRESSURE AND HEART RATEThe Lancet, 1983
- The Indirect Method of Recording Blood PressureCardiovascular Research, 1968