Malaria Rapid Diagnostic Devices: Performance Characteristics of the Para Sight F Device Determined in a Multisite Field Study
Open Access
- 1 August 2001
- journal article
- research article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 39 (8) , 2884-2890
- https://doi.org/10.1128/jcm.39.8.2884-2890.2001
Abstract
Microscopic detection of parasites has been the reference standard for malaria diagnosis for decades. However, difficulty in maintaining required technical skills and infrastructure has spurred the development of several nonmicroscopic malaria rapid diagnostic devices based on the detection of malaria parasite antigen in whole blood. TheParaSight F test is one such device. It detects the presence of Plasmodium falciparum-specific histidine-rich protein 2 by using an antigen-capture immunochromatographic strip format. The present study was conducted at outpatient malaria clinics in Iquitos, Peru, and Maesod, Thailand. Duplicate, blinded, expert microscopy was employed as the reference standard for evaluating device performance. Of 2,988 eligible patients, microscopy showed that 547 (18%) had P. falciparum, 658 (22%) had P. vivax, 2 (0.07%) had P. malariae, and 1,750 (59%) were negative for Plasmodium. Mixed infections (P. falciparum and P. vivax) were identified in 31 patients (1%). The overall sensitivity of ParaSight F forP. falciparum was 95%. When stratified by magnitude of parasitemia (no. of asexual parasites per microliter of whole blood), sensitivities were 83% (>0 to 500 parasites/μl), 87% (501 to 1,000/μl), 98% (1,001 to 5,000/μl), and 98% (>5,000/μl). Device specificity was 86%.Keywords
This publication has 40 references indexed in Scilit:
- Use of Dipstick Tests for the Rapid Diagnosis of Malaria in Nonimmune TravelersJournal of Travel Medicine, 2006
- ParaSight™-F test results in cerebral malaria patients before and after treatment in Chittagong Medical College Hospital, BangladeshTransactions of the Royal Society of Tropical Medicine and Hygiene, 2000
- Reliability of malaria microscopy in epidemiological studies: results of quality controlTropical Medicine & International Health, 2000
- Comparison of a rapid field immunochromatographic test to expert microscopy for the detection of Plasmodium falciparum asexual parasitemia in ThailandActa Tropica, 1999
- An assessment of the accuracy of clinical diagnosis, local microscopy and a rapid immunochromatographic card test in comparison with expert microscopy in the diagnosis of malaria in rural KenyaTransactions of the Royal Society of Tropical Medicine and Hygiene, 1999
- ACP Broadsheet no 148. July 1996. Laboratory diagnosis of malaria.Journal of Clinical Pathology, 1996
- Accuracy of routine laboratory diagnosis of malaria in the United Kingdom.Journal of Clinical Pathology, 1994
- Diagnosis of malaria by detection of Plasmodium falciparum HRP-2 antigen with a rapid dipstick antigen-capture assayThe Lancet, 1994
- The rapid manual ParaSight®-F test. A new diagnostic tool for Plasmodium falciparum infectionTransactions of the Royal Society of Tropical Medicine and Hygiene, 1993
- Malaria epidemiology and detectabilityTransactions of the Royal Society of Tropical Medicine and Hygiene, 1982