Detection of Chlamydia trachomatis in vaginal specimens from female commercial sex workers using a new improved enzyme immunoassay
Open Access
- 1 December 1998
- journal article
- Published by BMJ in Sexually Transmitted Infections
- Vol. 74 (6) , 435-438
- https://doi.org/10.1136/sti.74.6.435
Abstract
OBJECTIVE: To evaluate the performance of a new improved enzyme immunoassay (EIA) kit for the detection of Chlamydia trachomatis in vaginal swab and endocervical swab specimens from female commercial sex workers, in comparison with a conventional EIA test and a polymerase chain reaction (PCR) assay. METHODS: A high risk group of 163 female commercial sex workers who visited a sexually transmitted disease (STD) clinic in order to undergo screening for major STDs, including chlamydial infection, were enrolled. A total of four swab specimens, including two vaginal and two endocervical specimens, were collected from each woman by a clinician. To identify C trachomatis, a new improved EIA kit (IDEIA PCE), a conventional EIA kit (IDEIA), and PCR assay (Amplicor) were used. Discrepancies in the results were resolved using supplementary PCR assay. A female patient was considered to be infected with C trachomatis if the IDEIA PCE test and PCR test for both sample sites (endocervical and vaginal) gave positive results. Following resolution of these discrepancies, relative sensitivity and specificity, confidence intervals, and predictive values for each type of specimen by each assay were calculated. RESULTS: Of the 163 women tested, 35 (21.5%) were shown to be infected with C trachomatis. The relative sensitivities in vaginal swab specimens were 88.8%, 68.6%, and 91.4% using IDEIA PCE, IDEIA, and PCR, respectively. The relative specificities in vaginal swab specimens were 99.2%, 99.2%, and 100%, respectively. The relative sensitivities in endocervical swab specimens were 85.7%, 77.1%, and 91.4% with IDEIA PCE, IDEIA, and PCR, respectively. The relative specificities in endocervical swab specimens were all 100%. CONCLUSIONS: The results obtained in this study suggest that the sensitivity and specificity of IDEIA PCE test on vaginal swab and endocervical swab specimens were similar to those of PCR assay on the two types of specimen. It is concluded that IDEIA PCE test on vaginal swab specimens is an acceptable, sensitive, and less invasive approach for the detection of C trachomatis in commercial sex workers with a high prevalence of C trachomatis infection.Keywords
This publication has 29 references indexed in Scilit:
- Basic Evaluation of Chlamydia Antigen Detection by EIA Using a Dual Amplification Enhanced Immunoassay MethodKansenshogaku Zasshi, 1998
- Diagnosis of urogenital Chlamydia trachomatis infection in women based on mailed samples obtained at home: multipractice comparative studyBMJ, 1996
- Trends in sexually transmitted diseases and condom use patterns among commercial sex workers in Fukuoka City, Japan 1990-93.Sexually Transmitted Infections, 1996
- The discrepancy in discrepant analysisThe Lancet, 1996
- Detection of Chlamydia trachomatis and Trichomonas vaginalis by polymerase chain reaction in introital specimens from pregnant womenAmerican Journal of Obstetrics and Gynecology, 1996
- Diagnosis of Chlamydia trachomatis genitourinary infection in women by ligase chain reaction assay of urineThe Lancet, 1995
- Diagnosis Of Chlamydia Trachomatis Urethral Infection In Symptomatic And Asymptomatic Men By Testing First-Void Urine In A Ligase Chain Reaction AssayThe Journal of Infectious Diseases, 1994
- Ligase chain reaction to detect Chlamydia trachomatis infection of the cervixJournal of Clinical Microbiology, 1994
- Detection of Chlamydia trachomatis in urine by polymerase chain reaction: Preliminary resultsCurrent Therapeutic Research, 1993
- Detection of Chlamydia trachomatis in endocervical specimens by polymerase chain reactionJournal of Clinical Microbiology, 1992