Modelling simple helically delivered dose distributions
- 23 March 2005
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 50 (7) , 1505-1517
- https://doi.org/10.1088/0031-9155/50/7/013
Abstract
In a previous paper, we described quality assurance procedures for Hi-Art helical tomotherapy machines. Here, we develop further some ideas discussed briefly in that paper. Simple helically generated dose distributions are modelled, and relationships between these dose distributions and underlying characteristics of Hi-Art treatment systems are elucidated. In particular, we describe the dependence of dose levels along the central axis of a cylinder aligned coaxially with a Hi-Art machine on fan beam width, couch velocity and helical delivery lengths. The impact on these dose levels of angular variations in gantry speed or output per linear accelerator pulse is also explored.Keywords
This publication has 6 references indexed in Scilit:
- Quality assurance of a helical tomotherapy machinePhysics in Medicine & Biology, 2004
- Clinical helical tomotherapy commissioning dosimetryMedical Physics, 2003
- Analyses and applications of single scan dose profiles in computed tomographyMedical Physics, 2003
- A new look at CT dose measurement: Beyond CTDIMedical Physics, 2003
- TomotherapySeminars in Radiation Oncology, 1999
- Tomotherapy: A new concept for the delivery of dynamic conformal radiotherapyMedical Physics, 1993