Implantable Cardioverter-Defibrillators and Pregnancy
- 4 November 1997
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation
- Vol. 96 (9) , 2808-2812
- https://doi.org/10.1161/01.cir.96.9.2808
Abstract
Background The purpose of this investigation was to evaluate the outcome of pregnancy in women with implantable cardioverter-defibrillators (ICDs). Methods and Results A multicenter retrospective analysis was performed on women with an ICD who became pregnant. Data were collected on 44 patients. The mean age and ejection fraction at the time of the initial ICD implant were 25.6±4.9 years and 49.8±9.7%, respectively. The mean follow-up after the ICD implant was 4.8±2.8 years. Forty-two women had abdominally implanted generators, and 2 had a prepectoral device. Thirty had epicardial lead sensing systems, and 14 had transvenous. Thirty-six (82%) experienced no complications, and 8 (18%) had a medical or device-related complication. The ICD-related problems included tenderness at the ICD pocket scar (2 patients), generator migration (1), and pericarditis secondary to the epicardial patches (1). Medical complications were pulmonary embolism (1), therapeutic abortion (1), worsening hyperthyroidism (1), congestive heart failure (1), and weight loss (1). Thirty-seven women delivered vaginally, and 7 underwent cesarean section. Thirty-nine babies were born healthy, 1 was stillborn, 2 were small for gestational age, 1 had transient hypoglycemia, and 1 woman had a therapeutic abortion unrelated to the ICD. During pregnancy, 33 women received no ICD therapy, 8 had 1 shock, 1 had 5 discharges, 1 had 11 shocks, and 1 had 5 shocks. The total number of shocks during pregnancy ranged from 0 to 11, with an average of 0.66±1.9 discharges. Five women had 7 additional pregnancies without an ICD shock. Conclusions The mere presence of an ICD should not defer a women from becoming pregnant unless she has an underlying structural cardiac disease that is considered a contraindication. Pregnancy does not increase the risk of major ICD-related complications or result in a high number of ICD discharges.Keywords
This publication has 8 references indexed in Scilit:
- Treatment of arrhythmias during pregnancyPublished by Elsevier ,2004
- Effects of pregnancy on first onset and symptoms of paroxysmal supraventricular tachycardiaThe American Journal of Cardiology, 1995
- Cardiac Disease in PregnancyCardiology in Review, 1995
- Treatment of cardiac arrhythmias during pregnancyProgress in Cardiovascular Diseases, 1993
- Pregnancy complicated by chronic cardiomyopathy and an automatic implantable cardioverter defibrillatorAmerican Journal of Obstetrics and Gynecology, 1992
- New-onset ventricular tachycardia during pregnancyAmerican Heart Journal, 1992
- Paroxysmal ventricular tachycardia with syncope during pregnancyThe American Journal of Cardiology, 1986
- Disorders of the heartbeat during pregnancyAmerican Journal of Obstetrics and Gynecology, 1956