An Implantable Intracardiac Accelerometer for Monitoring Myocardial Contractility
- 1 December 1996
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 19 (12) , 2066-2071
- https://doi.org/10.1111/j.1540-8159.1996.tb03280.x
Abstract
As the myocardium contracts isometrically, it generates vibrations that are transmitted throughout the heart. These vibrations can be measured with an implantable microaccelerometer located inside the tip of an otherwise conventional unipolar pacing lead. These vibrations are, in their audible component, responsible for the first heart sound. The aim of this study was to evaluate, in man, the clinical feasibility and reliability of intracavity sampling of Peak Endocardial Acceleration (PEA) of the first heart sound vibrations using an implantable tip mounted accelerometer. We used a unidirectional accelerometer located inside the stimulating tip of a standard unipolar pacing lead: the sensor has a frequency response of DC to 1 kHz and a sensitivity of 5 mV/G (G - 9.81 m/s−2). The lead was connected to an external signal amplifier with a frequency range of 0.05–1,000 Hz and to a peak-to-peak detector synchronized with the endocardial R wave scanning the isovolumetric contraction phase. Following standard electro-physiological studies, sensor equipped leads were temporarily inserted in the RV of 15 patients (68 ± 15 years), with normal regional and global ventricular function, to record PEA at rest, during AAI pacing, during VVI pacing, and during dobutamine infusion (up to 20 |mg/kg per min). PEA at baseline was 1.1 G ± 0.5 (heart rate = 75 ± 14 beats/rain) and increased to 1.3 G ± 0.9 (P = NS vs baseline) during AAI pacing (heart rate = 140 beats/min) and to 1.4 G ± 0.5 (P = NS vs baseline) during VVI pacing (heart rate = 140 beats/min). Dobutamine infusion increased PEA to 3.7 G ± 1.1 (P < 0.001 vs baseline), with a heart rate of 121 ± 13 beats/min. In a subset of three patients, simultaneous hemodynamic RV monitoring was performed to obtain RV dP/dtmax, whose changes during dobutamine and pacing were linearly related to changes in PEA (r = 0.9; P < 0.001). In conclusion, the PEA recording can be consistently and safely obtained with an implantable device. Pharmacological inotropic stimulation, but not pacing induced chronotropic stimulation, increases PEA amplitude, in keeping with experimental studies, suggesting that PEA is an index ofmyocardial contractility. Acute variations in PEA are closely paralleled by changes in R V dP/dtmax, but are mainly determined by LV events. The clinical applicability of the method using RV endocardial leads and an implantable device offers potential for diagnostic applications in the long-term monitoring of myocardial function in man.Keywords
This publication has 13 references indexed in Scilit:
- A new method for measuring small local vibrations in the heart using ultrasoundIEEE Transactions on Biomedical Engineering, 1993
- Description of the deformation of the left ventricle by a kinematic modelJournal of Biomechanics, 1992
- Time-frequency transforms: a new approach to first heart sound frequency dynamicsIEEE Transactions on Biomedical Engineering, 1992
- Characterization of Natural and Total Artificial Heart AccelerationAnnual Northeast Bioengineering Conference, 1989
- Temporal Relation of the First Heart Sound to Closure of the Mitral ValveCirculation, 1973
- The Clinical Value of Frequency Analysis of the First Heart Sound in Myocardial InfarctionCirculation, 1970
- Usefulness and limitations of the rate of rise of intraventricular pressure (dp/dt) in the evaluation of myocardial contractility in man∗The American Journal of Cardiology, 1969
- Effects of Changing Heart Rate in Man by Electrical Stimulation of the Right AtriumCirculation, 1965
- STUDIES ON THE FIRST DERIVATIVE OF THE VENTRICULAR PRESSURE PULSE IN MANJournal of Clinical Investigation, 1962
- A Study of Heart Sounds and Murmurs by Direct Heart RecordingsCirculation, 1956