Evaluation of a new continuous thermodilution cardiac output monitor in critically ill patients
- 1 May 1995
- journal article
- research article
- Published by Wolters Kluwer Health in Critical Care Medicine
- Vol. 23 (5) , 860-866
- https://doi.org/10.1097/00003246-199505000-00014
Abstract
To evaluate the accuracy of a new continuous cardiac output monitor (one based on the thermodilution principle) in critically ill patients. Criterion standard study. Multidisciplinary intensive care unit in a university hospital. Fourteen critically ill patients, with different diseases, requiring pulmonary artery catheterization. In two patients with a left ventricular assist system, a defined, sudden 1 L/min change in cardiac output was carried through to evaluate the in vivo response time of the continuous cardiac output monitoring system. In the remaining 12 patients, cardiac output was altered by varying the dose of catecholamines, by volume loading, or by varying the level of sedation. In four patients, a rapid infusion of cold saline was given through a central venous catheter to test the performance of the system under these conditions. Cardiac output was monitored continuously. A total of 163 (13 to 18 per patient) bolus determinations of cardiac output were performed, using the conventional thermodilution technique and simultaneously using the indocyanine green dye dilution technique. The range of cardiac output was 3.8 to 15.6 L/min. The results of the continuous thermodilution method were compared with the results of the bolus thermodilution and the dye dilution methods, respectively. The mean difference (bias) +/- SD of differences (precision) was 0.35 +/- 1.01 L/min for continuous vs. bolus thermodilution and 0.34 +/- 1.01 L/min for continuous thermodilution vs. indocyanine green dye dilution. Linear regression (correlation) analyses were y = 0.95x + 0.76 (r2 = .91) for continuous and bolus thermodilution and y = 0.93x + 0.87 (r2 = .91) for continuous thermodilution and dye dilution. The 75% in vivo response time was 10.5 mins. The infusion of cold isotonic saline led to erroneous continuous cardiac output values. When the conventional bolus thermodilution and dye dilution techniques were compared, mean difference was -0.01 +/- 0.54 L/min and the results of linear regression analyses were y = 0.97x + 0.22 (r2 = .97). Continuous cardiac output measurement using the thermodilution technique is reasonably accurate and is reliable and applicable in routine clinical practice, and therefore may add to patient safety. However, the response time is too slow for the immediate detection of acute changes in cardiac output. Some clinical conditions such as the rapid infusion of cold solutions can interfere with the continuous cardiac output measurement. Conventional bolus thermodilution and indocyanine green dye dilution methods showed good agreement and can be used interchangeably.Keywords
This publication has 23 references indexed in Scilit:
- Thoracic electrical bioimpedance measurement of cardiac output in postaortocoronary bypass patientsCritical Care Medicine, 1993
- COMPARISON OF THORACIC ELECTRICAL BIOIMPEDANCE AND THERMODILUTION FOR THE MEASUREMENT OF CARDIAC INDEX IN PATIENTS WITH SEVERE SEPSISBritish Journal of Anaesthesia, 1993
- Continuous thermodilution cardiac output measurement in intensive care unit patientsJournal of Cardiothoracic and Vascular Anesthesia, 1992
- Continuous cardiac output measurements in critically III patientsJournal of Cardiothoracic and Vascular Anesthesia, 1992
- Simultaneous Intraoperative Measurement of Cardiac Output by Thermodilution and Transtracheal DopplerAnesthesiology, 1991
- Continuous measurement of cardiac output with the use of stochastic system identification techniquesJournal of Clinical Monitoring and Computing, 1990
- Simultaneous measurements of cardiac output by thermodilution, esophageal doppler, and electrical impedance in anesthetized patientsJournal of Cardiothoracic Anesthesia, 1988
- Pulsed Doppler echocardiographic determination of stroke volume and cardiac output: clinical validation of two new methods using the apical window.Circulation, 1984
- Thermodilution cardiac output determination with a single flow-directed catheterAmerican Heart Journal, 1972
- Measurement of cardiac output by thermal dilution in man.Journal of Applied Physiology, 1968