A comparison of the performance of 20 pulse oximeters under conditions of poor perfusion
- 1 January 1991
- journal article
- research article
- Published by Wiley in Anaesthesia
- Vol. 46 (1) , 3-10
- https://doi.org/10.1111/j.1365-2044.1991.tb09303.x
Abstract
Summary: The performance of 20 pulse oximeters with finger probes was evaluated by comparison of their readings with directly measured arterial blood oxygen saturations. The samples were taken from patients who had undergone cardiac surgery under hypothermic cardiopulmonary bypass and had poor peripheral perfusion. The mean difference (bias, accuracy), standard deviation (precision) and drop‐out rate for each pulse oximeter was determined. An overall ranking of performance of each pulse oximeter was calculated using five criteria (accuracy, precision, number of readings within 3% of standard, percentage of readings given within 3% of standard, expectedoverread limit in 95% of cases). Two pulse oximeters achieved a combination of accuracy and precision such that 95% of measurements would be expected to be within 4% of the co‐oximeter value; these two also had the lowest dropout rate.Keywords
This publication has 13 references indexed in Scilit:
- Potential errors in pulse oximetryAnaesthesia, 1991
- COMPARISON OF PULSE OXIMETERS: EFFECTS OF VASOCONSTRICTION AND VENOUS ENGORGEMENTBritish Journal of Anaesthesia, 1989
- II. PULSE OXIMETERS: ESSENTIAL MONITORS WITH LIMITATIONSBritish Journal of Anaesthesia, 1989
- Principles of Pulse OximetryAnesthesia & Analgesia, 1989
- Peripheral perfusion after cardiac surgeryCritical Care Medicine, 1989
- Pulse oximetryJournal of Clinical Monitoring and Computing, 1989
- Pulse OximetryAnesthesiology, 1989
- ACCURACY OF A PULSE OXIMETER IN THE CRITICALLY ILL ADULTAnesthesiology, 1985
- Statistics and ethics in medical research. VI--Presentation of results.BMJ, 1980