20-, 23-, AND 25-GAUGE VITREOUS CUTTERS
- 1 February 2008
- journal article
- research article
- Published by Wolters Kluwer Health in Retina
- Vol. 28 (2) , 249-257
- https://doi.org/10.1097/iae.0b013e31815ec2b3
Abstract
The flow rate of nine vitreous cutters was tested in two different viscosity environments. Physical features of each probe, such as flexibility and internal lumen diameter, were measured. For each probe, a total of 26 aspiration tests were performed in two different liquids. The stiffness of the shaft of each probe was calculated by measuring the displacement of the tip of the probe under a known constant force. All the probes were dissected and the internal lumen diameter was measured. Significant differences in flow rates were observed among the three separate conditions, among the three different gauge groups (25-, 23-, and 20-gauge vitreous cutter), and also within each gauge group. Internal lumen diameters and probe stiffness characteristics are presented. Vacuum, internal lumen diameter, and duty cycle seem to be the critical features governing actual flow rate. Knowledge of flow rate and stiffness of the various probes may aid surgeons in making an educated choice and may improve patient care.Keywords
This publication has 8 references indexed in Scilit:
- 25-gauge vs 20-gauge system for pars plana vitrectomy: a prospective randomised clinical trialBritish Journal of Ophthalmology, 2007
- Comparaison des différents systèmes de vitrectomieJournal Français d'Ophtalmologie, 2005
- Longer-term Outcomes of Transconjunctival Sutureless 25-gauge VitrectomyAmerican Journal of Ophthalmology, 2005
- TRANSCONJUNCTIVAL SUTURELESS 23-GAUGE VITRECTOMYRetina, 2005
- A new 25-gauge instrument system for transconjunctival sutureless vitrectomy surgery1 1The new Transconjunctival Sutureless Vitrectomy System is disclosed to Bausch and Lomb Surgical, St. Louis, MO. The Microsurgery Advanced Design Laboratory (MADLAB) may receive royalties related to the sale of this and other instruments mentioned in the article.Ophthalmology, 2002
- Initial experience using the transconjunctival sutureless vitrectomy system for vitreoretinal surgery1 1The new Transconjunctival Sutureless Vitrectomy System is disclosed to Bausch & Lomb Surgical, St. Louis, MO. The Microsurgery Advanced Design Laboratory (MADLAB) may receive royalties related to the sale of this and other instruments mentioned in the article.Ophthalmology, 2002
- Changes in density and viscosity of chicken egg albumen and yolk during incubationJournal of Experimental Zoology, 1990
- An Introduction to Fluid Dynamics.The American Mathematical Monthly, 1969