Experimental results on fast 2D-encoded magnetic particle imaging
- 22 February 2008
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 53 (6) , N81-N84
- https://doi.org/10.1088/0031-9155/53/6/n01
Abstract
This paper presents the first experimental results on magnetic particle imaging with full 2D encoding. The encoding speed achieved was 3.88 ms for a field of view of 1x1 cm2. Small phantoms composed of several dots each filled with 200 nl undiluted Resovist (500 mmol(Fe) l(-1)) were scanned. A resolution of better than 1 mm was achieved for a frame rate of 25 frames s(-1).Keywords
This publication has 4 references indexed in Scilit:
- A simulation study on the resolution and sensitivity of magnetic particle imagingPhysics in Medicine & Biology, 2007
- Tomographic imaging using the nonlinear response of magnetic particlesNature, 2005
- Limits of detection of SPIO at 3.0 T using T2* relaxometryMagnetic Resonance in Medicine, 2005
- Magnetic Resonance ImagingPublished by Springer Nature ,2003