ACCEPTABLE AGREEMENT BETWEEN TYMPANIC AND RECTAL TEMPERATURE IN ACUTE STROKE PATIENTS
- 1 March 2002
- journal article
- Published by Wiley in International Journal Of Clinical Practice
- Vol. 56 (2) , 82-84
- https://doi.org/10.1111/j.1742-1241.2002.tb11201.x
Abstract
Major concerns remain about the reliability of tympanic thermometers. The aim of this study was to investigate the agreement between rectal and tympanic thermometry and compare this with the repeatability of each method in patients with acute stroke. Tympanic temperature readings were from 0.7 degrees C below to 0.8 degrees C above rectal readings. As to repeatability, we found second rectal readings to be between 0.3 degrees C below and 0.4 degrees C above first readings, and second tympanic readings to be between 0.4 degrees C below and 0.5 degrees C above first readings. The agreement between the two methods is acceptable in clinical practice. Tympanic thermometry is useful in serial measurements, whereas, for a single accurate measurement, rectal mercury thermometry remains the method of choice.Keywords
This publication has 5 references indexed in Scilit:
- A comparison of four infrared tympanic thermometers with tympanic membrane temperatures measured by thermocouplesCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1996
- Screening for Fever in an Adult Emergency Department: Oral vs Tympanic ThermometrySouthern Medical Journal, 1996
- Re-evaluation of the tympanic thermometer in the emergency departmentAnnals of Emergency Medicine, 1992
- Infrared tympanic thermometerCritical Care Medicine, 1988
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986