A microbiological survey of automated biochemical machines.
Open Access
- 1 June 1980
- journal article
- research article
- Published by BMJ in Journal of Clinical Pathology
- Vol. 33 (6) , 566-570
- https://doi.org/10.1136/jcp.33.6.566
Abstract
The potential microbial hazards to laboratory personnel from operating three pieces of automated laboratory equipment have been tested using challenges of bacteria in serum or water. Aerosols did not seem to be a hazard, and the recommended disinfection procedures seemed quite adequate, although 2% aqueous activated glutaraldehyde is to be preferred to hypochlorite. The use of plastic gloves for protecting hands when changing rinse cups and pH electrodes is recommended.This publication has 7 references indexed in Scilit: