Fundamentals of Traveling Wave Ion Mobility Spectrometry
Top Cited Papers
- 6 November 2008
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 80 (24) , 9689-9699
- https://doi.org/10.1021/ac8016295
Abstract
Traveling wave ion mobility spectrometry (TW IMS) is a new IMS method implemented in the Synapt IMS/mass spectrometry system (Waters). Despite its wide adoption, the foundations of TW IMS were only qualitatively understood and factors governing the ion transit time (the separation parameter) and resolution remained murky. Here we develop the theory of TW IMS using derivations and ion dynamics simulations. The key parameter is the ratio (c) of ion drift velocity at the steepest wave slope to wave speed. At low c, the ion transit velocity is proportional to the squares of mobility (K) and electric field intensity (E), as opposed to linear scaling in drift tube (DT) IMS and differential mobility analyzers. At higher c, the scaling deviates from quadratic in a way controlled by the waveform profile, becoming more gradual with the ideal triangular profile but first steeper and then more gradual for realistic profiles with variable E. At highest c, the transit velocity asymptotically approaches the wave speed. Unlike with DT IMS, the resolving power of TW IMS depends on mobility, scaling as K1/2 in the low-c limit and less at higher c. A nonlinear dependence of the transit time on mobility means that the true resolving power of TW IMS differs from that indicated by the spectrum. A near-optimum resolution is achievable over an ∼300−400% range of mobilities. The major predicted trends are in agreement with TW IMS measurements for peptide ions as a function of mobility, wave amplitude, and gas pressure. The issues of proper TW IMS calibration and ion distortion by field heating are also discussed. The new quantitative understanding of TW IMS separations allows rational optimization of instrument design and operation and improved spectral calibration.Keywords
This publication has 52 references indexed in Scilit:
- Control of Ion Distortion in Field Asymmetric Waveform Ion Mobility Spectrometry via Variation of Dispersion Field and Gas TemperatureAnalytical Chemistry, 2008
- Gas-Phase Proton-Transfer Chemistry Coupled with TOF Mass Spectrometry and Ion Mobility-MS for the Facile Analysis of Poly(ethylene glycols) and PEGylated Polypeptide ConjugatesAnalytical Chemistry, 2008
- Monitoring copopulated conformational states during protein folding events using electrospray ionization-ion mobility spectrometry-mass spectrometryJournal of the American Society for Mass Spectrometry, 2007
- Multiplexed Ion Mobility Spectrometry-Orthogonal Time-of-Flight Mass SpectrometryAnalytical Chemistry, 2007
- High-Resolution Field Asymmetric Waveform Ion Mobility Spectrometry Using New Planar Geometry AnalyzersAnalytical Chemistry, 2006
- Characterizing the Structures and Folding of Free Proteins Using 2-D Gas-Phase Separations: Observation of Multiple Unfolded ConformersAnalytical Chemistry, 2006
- Mapping the Proteome ofDrosophilamelanogaster: Analysis of Embryos and Adult Heads by LC−IMS−MS MethodsJournal of Proteome Research, 2005
- Applications of a travelling wave‐based radio‐frequency‐only stacked ring ion guideRapid Communications in Mass Spectrometry, 2004
- Electrospray ionization of large multiply charged species proceeds via Dole’s charged residue mechanismPublished by Elsevier ,2000
- Disulfide-Intact and -Reduced Lysozyme in the Gas Phase: Conformations and Pathways of Folding and UnfoldingThe Journal of Physical Chemistry B, 1997