A comparison of capillary electrophoresis using indirect UV absorbance and ICP-MS detection with a self-aspirating nebulizer interface
- 27 April 2000
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in The Analyst
- Vol. 125 (5) , 819-823
- https://doi.org/10.1039/a910268j
Abstract
A comparison of capillary electrophoresis (CE) migration times using standard on-column UV detection and inductively coupled plasma mass spectrometry (ICP-MS) is presented for two different CE separations. The first is a separation of five arsenic compounds, four anionic species and one neutral, using reverse polarity in a chromate buffer with an electroosmotic flow modifier. The second CE separation employs normal polarity for the separation of eleven lanthanide cations using a hydroxyisobutyric acid (HIBA) buffer with an indirect UV detection reagent. The CE-ICP-MS interface is based on a tee design which introduces a sheath flow around the CE capillary using a self-aspirating make-up electrolyte with a microconcentric nebulizer (MCN-100). Migration times observed for CE-ICP-MS electropherograms are comparable to migration times acquired with UV detection, and may be adjusted by changes in the make-up liquid level position. The precision of peak areas (−1 range for the analytes studied.Keywords