The Welch Allyn Audioscope and Microtymp: their accuracy and that of pneumatic otoscopy, tympanometry and pure tone audiometry as predictors of otitis media with effusion
- 1 July 1992
- journal article
- research article
- Published by Cambridge University Press (CUP) in The Journal of Laryngology & Otology
- Vol. 106 (7) , 600-602
- https://doi.org/10.1017/s0022215100120298
Abstract
Results of standard audiometry and tympanometry were compared with the Welch Allyn ‘Audioscope’ and ‘Microtymp’, in the diagnosis of secretory otitis media, in 100 children.Standard pure tone audiometry had a specificity of 92 per cent and sensitivity of 51.6 per cent. The Welch Allyn ‘Audioscope’ a specificity of 84.2 per cent and sensitivity of 57.5 per cent; standard tympanometry a specificity of 71 per cent and sensitivity of 88 per cent; the Welch Allyn ‘Microtymp’ a specificity of 63 per cent and sensitivity of 90 per cent.In view of cost, portability, and speed of testing the Welch Allyn instruments would be particularly suitable for community screening.Pure tone audiometry with a pass threshold of 25 dB, at 2 and 4 kHz is a poor indicator of effusion.Keywords
This publication has 6 references indexed in Scilit:
- Pneumatic otoscopy and tympanometry in the detection of middle ear effusionClinical Otolaryngology, 1990
- Predictive Value of Tympanometry in Middle Ear EffusionAnnals of Otology, Rhinology & Laryngology, 1986
- Identification of Otitis Media with Effusion in ChildrenAnnals of Otology, Rhinology & Laryngology, 1980
- The value of impedance testing in diagnosis of middle ear effusionThe Journal of Laryngology & Otology, 1976
- Audiometry and tympanometry in relation to middle ear effusions in childrenThe Laryngoscope, 1973
- Clinical Experience With Impedance AudiometryJAMA Otolaryngology–Head & Neck Surgery, 1970