Bioluminescent Flow Sensors: L-Alanine Determination in Serum and Urine
- 1 August 1987
- journal article
- research article
- Published by Taylor & Francis in Analytical Letters
- Vol. 20 (8) , 1315-1330
- https://doi.org/10.1080/00032718708064611
Abstract
A continuous flow bioluminescent method for L-alanine analysis in serum and urine has been developed. Serum can be analyzed directly after simple filtration. Response is linear from 50 to 1500 pmoles in biological matrix. Alanine dehydrogenase is immobilized onto a nylon coil separated from the reactor coil containing bioluminescent enzymes. The stability of nylon immobilized enzymes is high (over three months) and more than 900 samples can be analyzed with few mg of enzymes. The results obtained with the bioluminescent sensor agree well with those obtained by ion exchange chromatography (amino acid analyzer).Keywords
This publication has 3 references indexed in Scilit:
- Bioluminescent continuous-flow assay of adenosine 5'-triphosphate using firefly luciferase immobilized on nylon tubesAnalytical Chemistry, 1986
- [21] Flow systems utilizing nylon-immobilized enzymesPublished by Elsevier ,1986
- Bioluminescent assays of picomole levels of various metabolites using immobilized enzymesAnalytical Biochemistry, 1982