Analytic Bias of Thyroid Function Tests: Analysis of a College of American Pathologists Fresh Frozen Serum Pool by 3900 Clinical Laboratories
Open Access
- 1 March 2005
- journal article
- Published by Archives of Pathology and Laboratory Medicine in Archives of Pathology & Laboratory Medicine
- Vol. 129 (3) , 310-317
- https://doi.org/10.5858/2005-129-310-abotft
Abstract
Context.—In proficiency testing surveys, there are differences in the values reported by users of various analytic methods. Two contributors to this variation are calibrator bias and matrix effects of proficiency testing materials. Objectives.—(1) To quantify the biases of the analytic methods used to measure thyroid-stimulating hormone, thyroxine, triiodothyronine, free thyroxine, and free triiodothyronine levels; (2) to determine if these biases are within allowable limits; and (3) to ascertain if proficiency testing materials correctly identify these biases. Design.—A fresh frozen serum specimen was mailed as part of the 2003 College of American Pathologists Ligand and Chemistry surveys. The means and SDs for each analytic method were determined for this sample as well as for a proficiency testing sample from both surveys. In the fresh frozen serum sample, target values for thyroxine and triiodothyronine were determined by isotope dilution/liquid chromatography/tandem mass spectrometry. All other target values in the study were the median of the means obtained for the various analytic methods. Main Outcome Measures.—Calibration biases were calculated by comparing the mean of each analytic method with the appropriate target values. These biases were evaluated against limits based on intra- and interindividual biological variation. Matrix effects of proficiency testing materials were assessed by comparing the rank of highest to lowest analytic method means (Spearman rank test) for each analyte. Participants.—Approximately 3900 clinical laboratories were enrolled in the College of American Pathologists Chemistry and Ligand surveys. Results.—The number of methods in the Ligand Survey that failed to meet the goals for bias was 7 of 17 for thyroid-stimulating hormone and 11 of 13 for free thyroxine. The failure rates were 12 of 16 methods for thyroxine, 8 of 11 for triiodothyronine, and 9 of 11 for free triiodothyronine. The means of the analytic method for the proficiency testing material correlated significantly (P < .05) only with the fresh frozen serum means for thyroxine and thyroid-stimulating hormone in the Chemistry Survey and free triiodothyronine in the Ligand Survey. Conclusions.—A majority of the methods used in thyroid function testing have biases that limit their clinical utility. Traditional proficiency testing materials do not adequately reflect these biases.Keywords
This publication has 11 references indexed in Scilit:
- Differences between measurements of T4 and T3 in pregnant and nonpregnant women using isotope dilution tandem mass spectrometry and immunoassays: are there clinical implications?Clinica Chimica Acta; International Journal of Clinical Chemistry, 2004
- Isotope dilution tandem mass spectrometric method for T4/T3Clinica Chimica Acta; International Journal of Clinical Chemistry, 2004
- Quality specifications in laboratory medicine - current consensus viewsAccreditation and Quality Assurance, 1999
- Commutability and traceability: their repercussions on analytical bias and inaccuracyClinica Chimica Acta; International Journal of Clinical Chemistry, 1999
- Isotope dilution-liquid chromatography/electrospray ionization-tandem mass spectrometry for the determination of serum thyroxine as a potential reference methodRapid Communications in Mass Spectrometry, 1998
- The evolution and limitations of accuracy and precision standardsClinica Chimica Acta; International Journal of Clinical Chemistry, 1997
- A conceptual model for establishing tolerance limits for analytic bias and imprecision based on variations in population test distributionsClinica Chimica Acta; International Journal of Clinical Chemistry, 1997
- Summary of the Second Report of the National Cholesterol Education Program (NCEP) Expert Panel on Detection, Evaluation, and Treatment of High Blood Cholesterol in Adults (Adult Treatment Panel II)JAMA, 1993
- Current status of blood cholesterol measurement in clinical laboratories in the United States: a report from the Laboratory Standardization Panel of the National Cholesterol Education Program.Clinical Chemistry, 1988
- Outliers in Statistical DataPhysics Today, 1979