Self‐Inhibition of an AV Sequential Demand (DVI) Pulse Generator Due to Polyurethane Lead Insulation Disruption
- 1 July 1983
- journal article
- case report
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 6 (4) , 840-844
- https://doi.org/10.1111/j.1540-8159.1983.tb04404.x
Abstract
Self-inhibition of bipolar DVI AV-sequential pacemakers has been attributed to electrode malposition, although the potential for self-inhibition due to current leakage has been recognized. Previous cases of current leakages due to insulation defects in bipolar VVI ventricular demand systems have usually presented as abnormal sensing or pacing function without a change to actual unipolar sensing and pacing. This article describes a case of disruption of the insulation of polyurethane bipolar atrial and ventricular leads by silk ligatures, resulting in unipolar atrial and ventricular pacing and self-inhibition of ventricular output of a DVI pulse generator. This case emphasizes the importance of avoiding any application of suture material directly to polyurethane leads. This case also emphasizes the importance of monitoring pacemaker artifact amplitude and axis in patient follow-up.Keywords
This publication has 9 references indexed in Scilit:
- Pseudo‐fracture of Pacemaker Lead Due to Securing Suture: A Case ReportPacing and Clinical Electrophysiology, 1981
- Autointerference of Demand Pulse GeneratorsPacing and Clinical Electrophysiology, 1981
- Apparent VVI (or ? VOO) Pacing by AV Sequential Demand (DVI) Puise GeneratorPacing and Clinical Electrophysiology, 1981
- Characterization of Pacemaker Arrhythmias Due to Normally Functioning AV Demand (DVI) Pulse GeneratorsPacing and Clinical Electrophysiology, 1980
- Transvenous Insertion of Double Sets of Permanent ElectrodesPublished by American Medical Association (AMA) ,1980
- Transvenous insertion of double sets of permanent electrodes. Atraumatic technique for atrial synchronous and atrioventricular sequential pacemakersJAMA, 1980
- Myopotential inhibition of a bipolar pacemaker caused by electrode insulation defect.Heart, 1977
- Suppression of a ventricular-inhibited bipolar pacemaker by skeletal muscle activityJournal of Electrocardiology, 1974
- Atrioventricular Sequential Pacing and PacemakersChest, 1973