On Stress Reduction in Bioprosthetic Heart Valve Leaflets by the Use of a Flexible Stent
- 1 December 1991
- journal article
- Published by Hindawi Limited in Journal of Cardiac Surgery
- Vol. 6 (4) , 476-481
- https://doi.org/10.1111/j.1540-8191.1991.tb00348.x
Abstract
It has been postulated that flexible stent posts can reduce tensile stress at the commissures of tissue heart valves by about 90% when compared with the same valve mounted on a rigid stent. We have used a detailed computer model to investigate the role of flexible stent posts in reducing stress in the leaflets of three types of bioprosthetic heart valves: the bovine pericardial and the high- and zero-pressure fixed porcine valves. The models use stress/strain data from biaxial experiments to characterize the tissue properties and are subjected to a back pressure of 120 mmHg. We found that strain was reduced linearly with stent post deflection and that this was a purely static process--it did not require the load to be applied impulsively. This finding was in close agreement with earlier experimental studies, which measured the same strain reduction whether the valve was loaded quasi-statically or at physiological rates. In addition, we found that for this mechanism to be effective the valve must have good coaptation at the center and the tissue should be stiff; in other cases, the advantages of strain reduction through the use of a flexible stent are considerably diminished.Keywords
This publication has 7 references indexed in Scilit:
- Identification of a Failure Mode of the Antibiotic Sterilized Aortic Allograft After 10 Years: Implications for Their Long-Term SurvivalJournal of Cardiac Surgery, 1991
- A mechanical analysis of the closed Hancock heart valve prosthesisJournal of Biomechanics, 1988
- Long-term followup of viable frozen aortic homograftsThe Journal of Thoracic and Cardiovascular Surgery, 1987
- Effect of glutaraldehyde fixation and valve constraint conditions on porcine aortic valve leaflet coaptation.Thorax, 1982
- Influence of fixation conditions on the performance of glutaraldehyde-treated porcine aortic valves: towards a more scientific basis.Thorax, 1979
- The glutaraldehyde-treated heterograft valveThe Journal of Thoracic and Cardiovascular Surgery, 1977
- The flexible stentThe Journal of Thoracic and Cardiovascular Surgery, 1971