The Control of Errors in Infrared Spectrophotometry. VI. The Evaluation of Optical Constants by Combined Transmission and Attenuated Total Reflection Measurements
- 1 November 1980
- journal article
- research article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 34 (6) , 652-656
- https://doi.org/10.1366/0003702804731050
Abstract
An improved technique is described for the determination of the optical constants of liquids in the infrared. It is based on a combination of transmission and attenuated total reflection (ATR) measurements. A novel application of the Kramers-Kronig transform function is involved whereby a single-valued integration constant is derived from a vector of refractive index measurements obtained by ATR. These are measured over a wavenumber range where the absorbance is low. It has been shown in earlier work that by an appropriate choice of a low refractive index contact material (in our case sodium chloride) the ATR method has high precision and accuracy under these conditions. The accuracy of the method is evaluated on the basis of a statistical treatment of the propagation of the estimated errors in the experimentally measured quantities, viz., the transmittance, the cell thickness, and the ATR measurements which establish the integration constant (anchor point) of the Kramers-Kronig transform function. The transmittance measurements are made at several cell thicknesses. The data reduction computer program, by which the optical constants are evaluated from the transmission measurements, monitors these sets of data and selects at each wavenumber the one having the minimal statistical error.Keywords
This publication has 7 references indexed in Scilit:
- Absolute Absorption Intensity and Dispersion Measurements on Some Organic Liquids in the InfraredApplied Spectroscopy, 1980
- The Measurement of the Optical Constants of Thin Solid Films in the InfraredApplied Spectroscopy, 1980
- A Reflection Spectrophotometer for the Measurement of the Optical Constants of Liquids in the InfraredApplied Spectroscopy, 1980
- The determination of the optical constants of benzene and chloroform in the i.r. by thin film transmissionSpectrochimica Acta Part A: Molecular Spectroscopy, 1976
- The control of errors in i.r. spectrophotometry—V. Assessment of errors in the evaluation of optical constants by transmission measurements on thin filmsSpectrochimica Acta Part A: Molecular Spectroscopy, 1976
- The control of errors in i.r. spectrophotometry—IV. Corrections for dispersion distortion and the evaluation of both optical constantsSpectrochimica Acta Part A: Molecular Spectroscopy, 1976
- The control of errors in i.r. spectrophotometry—III. Transmission measurements using thin cellsSpectrochimica Acta Part A: Molecular Spectroscopy, 1976