Gas-Liquid Chromatographic-Thermal Energy Analyzer Determination of N-Nitrosodimethylamine in Beer: Collaborative Study
- 1 May 1982
- journal article
- research article
- Published by Oxford University Press (OUP) in Journal of AOAC INTERNATIONAL
- Vol. 65 (3) , 720-729
- https://doi.org/10.1093/jaoac/65.3.720
Abstract
The GLC/TEA method for N-nitrosodimethylamine (NDMA) in beer was studied collaboratively by 13 laboratories from 7 countries. Collaborators were asked to analyze a total of 10 randomly labeled samples of beer consisting of the following duplicates: a naturally contaminated commercial beer; a beer extremely low (ca 0.1 ppb) in NDMA; and the low NDMA beer spiked with 0.5,1.9, and 5.0 ppb NDMA. The pooled repeatability and reproducibility coefficients of variation (CV) for all samples were 17% and 27%, respectively. However, when data from 2 laboratories (outliers) were omitted, the corresponding CV values improved considerably (11% and 15%, respectively). Variance analysis showed the presence of a significant laboratory-sample interaction when all data were used for analysis, but this interaction disappeared when data from the 2 outlying laboratories were excluded. The pooled percent recovery of the overall method (omitting outliers) was 101.4 ± 3.5. All the laboratories detected NDMA in the low NDMA beer. The method was adopted official first action.This publication has 6 references indexed in Scilit:
- Gas-Liquid Chromatographic-Thermal Energy Analyzer Determination of N-Nitrosodimethylamine in Beer at Low Parts per Billion LevelJournal of AOAC INTERNATIONAL, 1981
- Nitrosamines in Malt and Malt BeveragesJournal of Food Science, 1981
- Quality Assurance in the Analysis of Foods for Trace ConstituentsJournal of AOAC INTERNATIONAL, 1980
- A Biometrics Invited Paper. Topics in Variance Component EstimationBiometrics, 1971
- Processing Data for OutliersPublished by JSTOR ,1953
- THE DISTRIBUTION OF THE LARGEST OF A SET OF ESTIMATED VARIANCES AS A FRACTION OF THEIR TOTALAnnals of Eugenics, 1941