Standoff Detection of High Explosive Materials at 50 Meters in Ambient Light Conditions Using a Small Raman Instrument
Top Cited Papers
Open Access
- 1 June 2005
- journal article
- research article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 59 (6) , 769-775
- https://doi.org/10.1366/0003702054280612
Abstract
We have designed and demonstrated a standoff Raman system for detecting high explosive materials at distances up to 50 meters in ambient light conditions. In the system, light is collected using an 8-in. Schmidt–Cassegrain telescope fiber-coupled to an f/1.8 spectrograph with a gated intensified charge-coupled device (ICCD) detector. A frequency-doubled Nd: YAG (532 nm) pulsed (10 Hz) laser is used as the excitation source for measuring remote spectra of samples containing up to 8% explosive materials. The explosives RDX, TNT, and PETN as well as nitrate- and chlorate-containing materials were used to evaluate the performance of the system with samples placed at distances of 27 and 50 meters. Laser power studies were performed to determine the effects of laser heating and photodegradation on the samples. Raman signal levels were found to increase linearly with increasing laser energy up to ∼3 × 106 W/cm2 for all samples except TNT, which showed some evidence of photo- or thermal degradation at higher laser power densities. Detector gate width studies showed that Raman spectra could be acquired in high levels of ambient light using a 10 microsecond gate width.Keywords
This publication has 37 references indexed in Scilit:
- Selective functionalisation of TNT for sensitive detection by SERRSElectronic supplementary information (ESI) available: full experimental details on the synthesis and analysis of the reported compounds. See http://www.rsc.org/suppdata/cc/b1/b110972c/Chemical Communications, 2002
- Interpretation of Raman Spectra of Nitro-Containing Explosive Materials. Part I: Group Frequency and Structural Class MembershipApplied Spectroscopy, 1997
- Raman spectroscopy for mineral identification and quantification for in situ planetary surface analysis: A point count methodJournal of Geophysical Research, 1997
- Fourier TransformRaman Spectroscopy of Some Energetic Materials and Propellant Formulations. IIJournal of Raman Spectroscopy, 1996
- Fourier transform Raman spectroscopy of some energetic materials and propellant formulationsJournal of Raman Spectroscopy, 1994
- Inexpensive, rapid field screening test for 2,4,6-trinitrotoluene in soilAnalytica Chimica Acta, 1992
- Membrane for in situ optical detection of organic nitro compounds based on fluorescence quenchingAnalytica Chimica Acta, 1990
- Single fiber absorption measurements for remote detection of 2,4,6-trinitrotolueneAnalytica Chimica Acta, 1989
- A clear, amine-containing poly(vinyl chloride) membrane for in situ optical detection of 2,4,6-trinitrotolueneAnalytica Chimica Acta, 1989
- Spectroscopic studies of explosives: 1—detection of nitro compounds on silica gel and carbon by non‐resonant Raman spectroscopyJournal of Raman Spectroscopy, 1983