Immunoturbidimetric assays for serum apolipoproteins A1 and B using Cobas Bio centrifugal analyser.
Open Access
- 1 April 1988
- journal article
- research article
- Published by BMJ in Journal of Clinical Pathology
- Vol. 41 (4) , 471-474
- https://doi.org/10.1136/jcp.41.4.471
Abstract
Immunoturbidimetric assays for measuring the apolipoproteins A1 and B using the Cobas Bio centrifugal analyser are described. The methods were specific, offered good sensitivity (less than 0.05 g/l) and intrabatch variability, with coefficients of variation between 2.4% and 3.5%, and were cost effective. Reference ranges were calculated for a group of civil servants, aged 35 to 55 years.Keywords
This publication has 7 references indexed in Scilit:
- Is the ratio of apo B/apo A-I an early predictor of coronary atherosclerosis?Atherosclerosis, 1986
- Plasma lipids and apolipoproteins as discriminators for presence and severity of angiographically defined coronary artery diseaseAtherosclerosis, 1983
- Standardization of immunoassays for the quantitation of plasma apo B proteinAnalytical Biochemistry, 1981
- Quantitative determination of human plasma apolipoprotein A-I by laser immunonephelometry.Clinical Chemistry, 1981
- Nephelometry of apolipoprotein B in human serum.Clinical Chemistry, 1979